Atta TASEH, Lercan ASLAN, Mohammad Hossein Chegini KORD, Mohamad AL MASRI, Ahmad Hedayatzadeh RAZAVI, Nazanin NAFISI, Colin O'NEILL, Kendal TOY, Joaquin BATISTA, Gregory WARYASZ, Christopher P. MILLER, Ara NAZARIAN, Soheil ASHKANI-ESFAHANI
Acta Orthopaedica et Traumatologica Turcica - 2026;60(5):1-7
Objective: This study aimed to compare the biomechanical properties of bio-inductive implant augmentation and suture repair for anterior talofibular ligament (ATFL) and calcaneofibular ligament (CFL) repair in chronic lateral ankle instability. Methods: This controlled cadaveric biomechanical study included eight fresh-frozen human ankle specimens [median age of 76 years, Interquartile Range (IQR): 70.5-83]; 5 males and 3 females). Using a repeated-measures design, each specimen was tested under three conditions: intact ligaments, suture repair, and bio-inductive implant augmentation (Bio-Aug). Following intact testing, the order of the suture repair and bio-inductive augmentation conditions was randomized. Biomechanical testing consisted of a 500-cycle loading protocol. The primary outcome measures were construct stiffness, primary creep, cyclic loading survival, and load-to-failure. Results: All specimens in the intact and Bio-Aug groups completed the 500-cycle protocol, whereas only three of eight (37.5%) specimens in the suture repair group did so. Both primary and final stiffness were greatest in the intact and Bio-Aug groups and lowest in the suture repair group; however, no significant differences were observed among groups for either primary (Friedman chi²(2)=4.33, P =.11) or final stiffness (Friedman chi²(2)=3.25, P =.20). Primary creep differed significantly across conditions (P=.006). Compared with the intact group, both the suture repair group (4.54% vs. 0.50%, P =.04) and the Bio-Aug group (1.60% vs. 0.50%, P =.02) demonstrated greater creep, although Bio-Aug values were substantially closer to those of the intact ligaments. The median load-to-failure of the Bio-Aug constructs was 330 N (IQR, 300-360 N). Conclusion: Bio-inductive implant augmentation restored construct stiffness comparable to that of the intact ligament complex while demonstrating substantially lower creep than suture repair alone in this time-zero cadaveric model. These findings suggest that bio-inductive augmentation may provide improved early mechanical support while preserving biomechanical characteristics closer to native ligament tissue. Further in vivo and clinical studies are required to determine whether these biomechanical advantages translate into improved clinical outcomes and rehabilitation protocols.