Huseyin TEPETAM, Cemal Ugur DURSUN, Sevecen Seyhun Nasir CAN, Mustafa Mert HANILCE, Gun GUNALP, Duygu GEDIK, Sule Karabulut GUL
İstanbul Kuzey Klinikleri Dergisi - 2026;13(5):544-550
OBJECTIVE: The objective of the study was to evaluate the agreement of set-up shift measurements obtained by the TomoTherapy automatic image-registration system, radiation therapists (RTTs), and physicians during adjuvant breast radiotherapy, and to investigate the effects of RTT number and target volume on positioning accuracy. METHODS: This retrospective single-center study included 40 female breast cancer patients treated with adjuvant helical TomoTherapy under daily megavoltage computed tomography (MVCT) guidance. Set-up shifts in the lateral, longitudinal, vertical, and roll directions were analyzed over 1,200 treatment fractions. Pairwise comparisons among the TomoTherapy system, RTTs, and physicians were performed using paired t-tests or Wilcoxon signed-rank tests with Bonferroni correction. The effect of RTT number was evaluated using the Wilcoxon signed-rank test within subgroups, and that of clinical target volume (CTV)/boost volume using the Kruskal-Wallis test. RESULTS: High agreement was observed among the TomoTherapy system, RTTs, and physicians. The largest mean difference between RTT and physician measurements was 1.56 mm and remained within the institutional 5-mm CTV-to-planning target volume margin. Although statistically significant differences were found in selected translational axes, RTT number and CTV/boost volume did not significantly influence set-up shift values. CONCLUSION: Daily MVCT-guided TomoTherapy provides reliable and reproducible patient positioning. Standardized RTT-led image-guided workflows maintain clinically acceptable set-up accuracy regardless of RTT team size or target volume.